A2463:Hires spectral scan of W51 with mocks

may08

History
Frequencies that were incorrect
W51 data for each "good" day

Misc
19may10: power of 2 jumps in spectra. testing
21may10: w51 on/off-1:
20may10: test decimations 1-20: power, voltage, jumps, and bandpass shape.
27jul10: jumps in xband spectra.


Intro:

    A2463 is doing a hi resolution spectral scan of W51 using the mock spectrometer in single pixel mode. To get the hi resolution, both mock groups are being used. As an example, the cband setup is:

History:


Freq that were incorrect  (top)

    For data taken 100518 thru 100526 some of the  data in the second group were taken at the wrong frequency.
When these data are plotted, the frequency axis is showing the expected value , not the value that was actually used. In the list below  the bands with bad frequency are listed. When it says bm3 2*df lo, the reported frequency is 237.464 MHz above the actual data recorded.



W51  data for each day (top)

    The w51 on/off''s were scaled to kelvins and then averaged for each day. The plots show:

processing: usr/a2463_xxmay10.pro


19may10: power of 2 jumps seen in spectra.

    Spectral jumps with power of 2 channels spacing were seen in the data.
The plots show various spectra with their jumps (.ps) (.pdf):
Summary:
processing: x101/100519/testa2463summary.pro

21may10: w51 on/off -1  (top)

    6 5min on/off-1 position switched patterns were computed and the averaged. The 2 pols were then also averaged.
The  plot shows the averaged spectra 4500 to 5050 (.ps) (.pdf)
processing: x101/100521/a2463.pro

27jul10:jumps in xband freq channels (top)

    On 27jul10 a2463 took data with xband receiver. The setup was:
    The jumps were along the freq axis. The trouble occurred:

files with problems 27jul10
fnum
bm
grp
freq
300
1
1
8237
400
0
1
8087
1100
1
1
9237
1200
0
1
9087
1500
5
1
9837
1900
1
1
9237
2800
0
1
9087
3100
0
1
9087
3500
0
1
9087
3500
5
1
9837

processing: usr/a2463/freqjumps/27jul10.pro