Intro
On 26jun22 crosses were done on casA with
the xband receiver. This was to check things after the
receiver was reinstalled following the QRFH testing. While
monitoring the data, a dip was seen in the total power trace
of polA. It lasted for a little more than 5 minutes. It
dropped about 4% and then returned to the original
value.
Setup and data taking
- cross pointing pattern:
- 10 sec cal On
- 10 sec cal Off
- 120 sec azimuth strip of .6 deg (great circle)
- 120 sec el strip of .6 deg
- Data taking:
- 7x172MHz bands with 1024 channels each
- 10hz sampling
Plotting the total power
The plots show the measured total
power (.ps) (.pdf) :
- Page 1 has data from the 8222 MHz band
- Page 2 had data from the 8504 MHz band
- Black is polA, red is polB
- The narrow positive spikes are the 10 second calOn
- This is followed by a 10 sec calOff , the azstrip over the
source, and then the el strip over the source.
- top Frame: Total power vs time.
- Data was normalized to the first 10sec caloff.
- Around 540 seconds polA starts to decrease, hits a
minimum, and the comes back up.
- the dip lasted for about 400 secs (actually a
little longer since i've left out the move time from
azstrip to el strip).
- There was a cal measurement at 740 seconds (during the
dip).
- Middle frame: blowup of the vertical axis.
- The blue line marks the minimum of the dip (.966 8222
band, .960 8504MHz band)
- Bottom frame: cal Deflection normalized to first calOn -
calOff
- The * mark the location and amplitude of the cal
deflection.
- The cal diode is temperature stabilized so its output
power should be constant.
Summary
The dip could be a gain change, a noise
level change, or both.
- with only a noise level change, the cal deflection should
remain constant
- A gain only variation should see the cal deflection change
by the same amount as the noise floor.
- the cal deflection changes by a little more than the
noise floor.
- Looking at the cal deflection during the dip, the noise
floor is increasing during the calOn,calOff. This causes
the cal deflection to be a little less than the noise
floor deflection.
- The cal deflections show that both polA and polB gains are
changing with time:
- polA: 2% in 1300 seconds
- polB: 3% in 1300 seconds
- these changes are probably electronic gain changes with
temperature.
- Not sure what would cause the 4% gain dip and then
recovery after 400 seconds.
- The dip occurred in multiple frequency bands:
- The problem occurs before the 2nd mixing stage if1 ->
if2
- It can be in the : lna,cables, postamp chassis, fiber
optics, or udc..
processing:x101/220624/220626_xbpolAdip.pro