Using the cal for a bandpass correction

nov 07, 2002

     The best baselines have historically been achieved by using position switching. You would track the source for 5 minutes and then spend 5 minutes off source retracing the same part of the dish. This  cancels the band shape from the IFLO and any ground based standing waves (standing waves from the source will not be canceled).  The down side of this method is that half the time is spent off source. This has typically been used for wide band extra galactic observations.

    Narrower band observations have tried observing with no off position. The bandpass can be removed by using a smoothed or fitted version of the on data. An alternative is to use the cal deflection to remove the bandpass created downstream from the cal injection. Let Txxx be the temperature from the xxx location and Fxxx be the frequency dependence of this temperature then:
 

    calOff=(((Tsrc*Fsrc + Tsky*Fsky1 + Tgrnd*Fgrnd1)*Fant1 + Trcv)*Fiflo

    calOn =(((Tsrc*Fsrc + Tsky*Fsky2 + Tgrnd*Fgrnd2)*Fant2 + Trcv + Tcal)*Fiflo
    Where 1,2 are taken at the calOff and calOn times. If these times are close together then 1=2 and:
        caldif=calon-caloff = Tcal*Filfo
 

    You could then use the calDif to divide out the bandpass from the IFLO. Any frequency dependence in front of the cal injection would remain. The example below use a 5 minute on, 5 minute off position switch on a galaxy with the lbn receiver. At the end of the off position, the cal is turned on for 10 seconds and then turned off for 10 seconds. The correlator setup was 9 level sampling with 2048 channels over 25 Mhz. The different page/figures show:

  • Fig 1: 5 minute on/off -1 bandpass
    5minute on/ 10secOff

  • calOn-calOff spectra
    5 minuteOn/(calon-caloff)
  • Fig 2:  smoothing (calOn-caloff) to increase signal to noise
  • Fig 3: Fittin polynomials to Calon-calOff
  • Fig 4: Fitting polynomials to CalOff

  •  

    The resulting band passes are:

        To improve the rms, the calDif spectra was smoothed prior to doing the division:
  • Fig 2 top: The calDif spectra was smoothed by 31 (.38 Mhz). This should compensate for the 300sec/10sec integration difference between the cal and the posOn. The red lines show a sine wave at the period of the smoothing. The rms has decreased from .0865 to .0202. This is an improvement of 4.3. The expected improvement of sqrt(31)=5.6
  • Fig 2 2nd: This has a smoothing of 101 channels (1.23 Mhz). The measured rms is .0157 compared to an expected value of .0086 ( .0865/sqrt(101)).
  • Fig 2 3rd: A smoothing of 201 channels (2.45 Mhz) with an rms of .0152 vs an expected rms of .006.
  • Fig 2 bottom: Has a smoothing of 301 (3.67 Mhz) with an rms of .0154 vs an expected rms of .005

  • Fitting polynomials to Calon-CalOff
     

  • Fig 3. This shows fitting polynomials (5,9,12,15 order) to the calon-caloff. The best rms of .0138 used a 5th order polynomial. This gave comparable results to smoothing by 2.45 Mhz.
  • Fig 4. The polynomial fitting was done on the calOff rather than the calOn-calOff. The best rms was .0028 which was 3.6 times larger than the full 5minOn/5minOff.
  •     The larger smoothing values are biased because the filter is extending into the region where the bandpass is overshooting. In spite of this the smoothing is not bringing the rms anywhere near the theoretical value. As soon as the smoothing width reaches the systematic errors  caused by the ripples in the bandpass, the rms improvement will start to degrade.

    Summary:

        The calon- calOff can be used to remove the bandpass shape of things downstream of the cal injection. The shorter integration time of the cal could be compensated by smoothing the calDif spectra or fitting polynomials. Polynomial fitting of calOn-calOff does a better job than smoothing of calOn-calOff. If you have an off position, then polynomial fitting of just the off is better than using the calon-calOff (but then you must worry about your source signal).

        If the cal deflection is less than the system temperature, then any errors in the cal spectra and any standing waves from in front of the cal injection will be multiplied by this factor when the calDeflection is normalized prior to the divide. With the present cal size, using the calDeflection does not appear to be practical. If we had a cal value that was many times the system temperature, and if the correlator linearity extended over this range, then any ripples in the calDifference spectra would be diminished rather than amplified when the calDifference spectra was normalized to unity prior to the division.

        Fitting polynomials to the calOff does a better job than using calOn-calOff. It would interesting to know if the extra structure from the calOn-calOff is just from the small cal, or there is frequency structure in the cal from say reflection of the cal signal going out the horn, reflecting off something and then coming back in.

        All of calDifference spectra will not do anything to the standing waves that are coming in from in front of the cal Injection point.
     
     

    processing: x101/obstech/calbpc/callbn.pro