Beam4a Tsys 10oct11

11oct11



Intro

Tsys measurement:

    On 10oct11 dana used the expensive spectrum analyzer to verify that the standard Tsys  measurements on bm4a were correct. He did this with the 12db amp inserted and then with the 16db amp. Tsys was taken with the mock spectrometers and then the spectrum analyzer (they were not taken simultaneously). The setups were:

The measurement results:

    The table below shows the tsys values measured. The last column shows the gains that the mock spectrometer used for beam 4a and beam 4b (a normal beam).



setup
TsysBm4a [K]
spec Analyzer
TsysBm4a [K]
mock Spectrometer
mockGain [db]
bm4a,bm4b
12db amp
72.1
71.5
52.9
18/11
16db amp
59.0
60.7
62.3
11/11


Plots were made showing the data used for the Tsys done with the mock spectrometer (.ps) (.pdf):

Summary:



processing: x101/111010/chkalfatsys.pro


home_~phil