alfa performance at OH (no 10db amps)
02apr05
links to plots:
comparison
of the system performance for alfa and lbw at OH (.ps) (.pdf)
A filter bank was installed in the alfa receiver
in mar05 that allowed switching between the 300 Mhz filters and filters
that let you look at the OH line. On 01apr05 the system m performance was
measured using:
-
1 minute on/off position switching
-
the source CTA21 (B0316+162. 7.2 Jy at 1666)
-
The data was centered at 1666 with a low side 1st lo.
-
The wapps were setup for 1 second dumps, 6.125 Mhz bandwidths.
-
on/offs were also done with lbw using the 1500 to 1800 Mhz filter.
Since the alfa
cals do not work at 1666 Mhz the metrics used were srcDeflection/Tsys
(G/T) and the ratio of G/T for alfa and lbw.
The plots show the comparison
of the system performance for alfa and lbw at OH (.ps) (.pdf)
-
Fig 1: Average on source and off source. Black is polA, red
is polB. The top plot is alfa while the bottom plot is lband wide. These
plots include the system temperature. The za's for the measurements were
alfa:7.1, lbw:11.5 degrees. The power levels were adjusted to the optimum
levels (1.) on source. The off source levels dropped to about .4 of optimum.
-
Fig 2: This shows srcDeflection/Offsrc. This is G/T. The two polarization's
of alfa differ by about 40%. This source is not polarized at lband. Using
the fluxes of 7.2 Jy, the sefd for the measurements are: Alfa: 4.3,6.0
Lbw: 2.4,2.3 Jy.
-
Fig 3: This shows (G/T alfa)/(G/T lbw) The average values are: .55
PolA, .37 polB.
The only question with these measurements are the low
values for the off source power. Lbw is probably ok since 3 db below the
optimum power level does not increase Tsys. I doubt that it should matter
with alfa either.
The original reason for the oh filters were for comet
observations that do a hexmap about the comet nucleus. With alfa
this could be done 7 times faster (if the outer beams were in the correct
position). The effectiveness is determined by the square of the ratios
of the sefd's. The numbers to use are:
-
sefdLbw: (2.4+2.3)/2 = 2.35
-
sefdAlfa: The outer pixels have an sefd about 1 db higher than the center
pixel (see sysperf
page). The averaged sefd would then be (4.3+6.0)/2. * (1
+6*1.2)/7. = 6.0
-
For alfa to match the sensitivity of lbw alfa needs to integrate (6.0/2.35)^2
= 6.5 times longer.
Lbw must also move between positions. For a1851 this was taking about 5%
longer. The relative sensitivity is:
-
lbw time: 7*1.05=7.4
-
Alfa time: 6.5
-
alfa is more sensitive than lbw by sqrt(7.4/6.5) = 1.06
The comparisons are then:
-
The sensitivities are about equal (alfa 1.06 time more sensitive).
-
Alfa allows simultaneous measurements. This is an advantage if things are
changing on the time frame of a hexmap pattern.
-
Lbw has the advantage that you can position the off positions anywhere
you want, while alfa restricts you to rotations at the beam spacing.
-
Calibration using the cals is available with lbw while it does not work
for alfa.
-
Polarization information is available with lbw but probably not with alfa.
processing: x101/050401/doit.pro