<< HP 8590D/E Product PageHewlett
Packard
8590D/E
|
Specifications apply to 8591E, 8593E, 8594E, 8595E, or 8596E unless otherwise noted. | ||||
FREQUENCY | ||||
Frequency Range | ||||
8591E | 50 Ω:
9 kHz to 1.8
GHz 75 Ω 1 MHz to 1.8 GHz |
|||
8594E | dc-coupled:
9 kHz to 2.9 GHz ac-coupled: 100 kHz to 2.9 GHz |
|||
8595E | dc-coupled:
9 kHz to 2.9 GHz ac-coupled: 100 kHz to 6.5 GHz |
|||
8596E | Band | LO harmonic = N | Center frequency | |
0 | 1 | 9 kHz to 2.9 GHz (dc-coupled) | ||
0 | 1 | 100 kHz to 2.9 GHz (ac-coupled) | ||
1 | 1 | 2.75 to 6.5 GHz | ||
2 | 2 | 6.0 to 12.8 GHz | ||
8593E | Band | LO harmonic = N | Center frequency | |
0 | 1 | 9 kHz to 2.9 GHz | ||
1 | 1 | 2.75 to 6.5 GHz | ||
2 | 2 | 6.0 to 12.8 GHz | ||
3 | 3 | 12.4 to 19.4 GHz | ||
4 | 4 | 19.1 to 22 GHz | ||
4 | 4 (Opt 026) | 19.1 to 26.5 GHz | ||
Frequency Reference | ||||
Aging | ± 2 X 10-7/year (Opt 004) | |||
Temperature Stability | ± 5 X 10-6; ± 1 X 10-8 (Opt 004) | |||
Initial Achievable Accuracy | ± 0.5 X 10-6; ± 2.2 X 10-8 (Opt 004) | |||
Frequency Readout Accuracy (start, stop, center, marker) | ± (frequency readout X freq ref error + span accuracy + 1% of span + 20% of RBW + 100 Hz X N) | |||
Marker Count Accuracy | ||||
Span ≤ 10 MHz X N | ± (marker freq X freq ref error + counter res + 100 Hz X N) | |||
Marker Count Accuracy | ||||
Span > 10 MHz X N | ± (marker freq X freq ref error + counter res + 1 kHz X N) | |||
Counter Resolution | ||||
Span ≤ 10 MHz X N | Selectable from 10 Hz to 100 kHz | |||
Span > 10 MHz X N | Selectable from 100 Hz to 100 kHz | |||
Frequency Span | ||||
Range | 0 Hz (zero span) and 8591E: 10 kHz to 1.8 GHz; 1 kHz min (Opt 130) 8594E: 10 kHz to 2.9 GHz; 1 kHz min (Opt 130) 8595E: 10 kHz to 6.5 GHz; 1 kHz min (Opt 130) 8596E: [10 X N] kHz to 12.8 GHz; [1 X N] kHz min (Opt 130) 8593E: [10 X N] kHz to 19.25 GHz; [1 X N] kHz min (Opt 130) |
|||
Resolution | Four digits or 20 Hz X N, whichever is greater | |||
Accuracy | ||||
Span ≤ 10 MHz X N | ± 2% of span | |||
Span > 10 MHz X N | ± 3% of span | |||
SWEEP TIME | ||||
Range | ||||
Span = 0 Hz or > 10 kHz | 20 ms to 100 s | |||
Span = 0 Hz (Opt 101) | 20 µs to 100 s | |||
Accuracy | ||||
20 ms to 100 s | ± 3% | |||
20 µs to < 20 ms (Opt 101) | ± 2% | |||
Sweep Trigger | free run, single, line, video, external | |||
Resolution Bandwidths | 1 kHz to 3 MHz (3 dB) in 1, 3, 10 sequence; 9 kHz and 120 kHz (6 dB) EMI bandwidths. Option 130 adds 30, 100, and 300 Hz (3 dB) bandwidths and 200 Hz (6 dB) EMI bandwidth. | |||
Accuracy | ± 20% | |||
Selectivity (characteristic) | -60 dB/-3 dB: 3 to 10 kHz, 15:1 100 kHz to 3 MHz, 15:1 1 kHz, 30 kHz, 16:1 -40 dB/-3 dB: 30 Hz to 300 Hz, 10:1 |
|||
Video Bandwidth Range | 30 Hz to 1 MHz in 1, 3 sequence (1 Hz to 1 MHz with Opt 130) | |||
STABILITY | ||||
Noise Sidebands (1 kHz RBW, 30 Hz VBW, sample detector) | ||||
> 10 kHz offset from CW signal | ≤-90 dBc/Hz + 20 log N | |||
> 20 kHz offset from CW signal | ≤-100 dBc/Hz + 20 log N | |||
>30 kHz offset from CW signal | ≤ -105 dBc/Hz + 20 log N | |||
Residual FM | ||||
HP 8591E | ||||
1 kHz RBW, 1 kHz VBW | ≤250 Hz p-p in 100 ms | |||
30 Hz RBW, 30 Hz VBW | ≤ 30 Hz p-p in 300 ms | |||
HP 8593E, 8594E, 8595E, 8596E | ||||
1 kHz, RBW, 1 kHz VBW | ≤ (250 X N) Hz p-p in 100 ms | |||
30 Hz RBW, 30 Hz VBW | ≤(30 X N) Hz p-p in 300 ms | |||
System Related Sidebands (> 30 kHz offset from CW signal) | ≤ -65 dBc + 20 log N | |||
Comb Generator (HP 8593E, 8596E) | 100 MHz fundamental frequency; ± 0.007% frequency accuracy | |||
AMPLITUDE | ||||
Amplitude Range | Displayed average noise level to +30 dBm | |||
HP 8591E Opt 001 | Displayed average noise level to +75 dBmV | |||
Maximum Safe Input (input attenuator @ 10 dB) | ||||
Average continuous power | +30 dBm (1Ω) | |||
HP 8591E Opt 001 | +75 dBmV (0.4 Ω) | |||
Peak Pulse Power | ||||
HP 8591E | +30 dBm (1 Ω) | |||
HP 8591E Opt 001 | +75 dBmV (0.4 Ω) | |||
HP 8593E, 8594E, 8595E, 8596E | +50 dBm (100 Ω) for < 10 µs pulse width and < 1% duty cycle, input atten > 30 dB | |||
DC | ||||
HP 8591E | 25 Vdc | |||
HP 8591E Opt 001 | 100 Vdc | |||
HP 8593E, 8594E, 8595E, 8596E | 0 V (dc-coupled); 50 V (ac-coupled) | |||
Gain Compression (> 10 MHz) | ≤0.5 dB (total power at input mixer = -10 dBm) | |||
Displayed Average Noise Level (input terminated, 0 dB atten, 30 Hz VBW or 1 Hz VBW with Opt 130, sample detector | ||||
30 Hz RBW | 1 kHz RBW | |||
HP 8591E | ||||
400 kHz to 1 MHz | ≤-130 dBm | ≤-115 dBm | ||
1 MHz to 1.5 GHz | ≤-130 dBm | ≤-115 dBm | ||
1.5 GHz to 1.8 GHz | ≤-128 dBm | ≤-113 dBm | ||
HP 8591E Opt 001 | ||||
1 MHz to 1.5 GHz | ≤-78 dBmV | ≤-63 dBm | ||
1.5 GHz to 1.8 GHz | ≤-76 dBmV | ≤-61 dBm | ||
HP 8594E | ||||
400 kHz to 5 MHz | ≤-122 dBm | ≤-107 dBm | ||
5 MHz to 2.9 GHz | ≤-127 dBm | ≤-112 dBm | ||
HP 8595E | ||||
400 kHz to 2.9 GHz | ≤-125 dBm | ≤-110 dBm | ||
2.75 to 6.5 GHz | ≤-127 dBm | ≤-112 dBm | ||
HP 8596E | ||||
400 kHz to 2.9 GHz | ≤-125 dBm | ≤-110 dBm | ||
2.75 to 6.5 GHz | ≤-127 dBm | ≤-112 dBm | ||
6.0 to 12.8 GHz | ≤-115 dBm | ≤-100 dBm | ||
HP 8593E | ||||
400 kHz to 2.9 GHz | ≤-127 dBm | ≤-112 dBm | ||
2.75 to 6.5 GHz | ≤-129 dBm | ≤-114 dBm | ||
6.0 to 12.8 GHz | ≤-117 dBm | ≤-102 dBm | ||
12.4 to 19.4 GHz | ≤-113 dBm | ≤-98 dBm | ||
19.1 to 22 GHz | ≤-107 dBm | ≤-92 dBm | ||
HP 8593E Opt 026 | ||||
19.1 to 26.5 GHz | ≤-102 dBm | ≤-87 dBm | ||
Spurious Responses | ||||
Second Harmonic Distortion | ||||
5 MHz to 1.8 GHz (HP 8591E) | < -70 dBc for -45 dBm tone at input mixer | |||
10 MHz to 2.9 GHz (HP 8593E) | < -70 dBc for -40 dBm tone at input mixer | |||
> 10 MHz (HP 8594E, 8595E, 8596E) | < -70 dBc for -40 dBm tone at input mixer | |||
> 2.75 GHz (HP 8593E, 8595E, 8596E) | < -100 dBc for two -30 dBm tones at input and > 50 kHz separation | |||
Third-order Intermodulation | ||||
HP 8591E (5 MHz to 1.8 GHz) | < -70 dBc for two -30 dBm tones at input and > 50 kHz separation | |||
HP 8593E, 8594E, 8595E, 8596E (> 10 MHz) | < -70 dBc for two -30 dBm tones at input and > 50 kHz separation | |||
Other Input-related Spurious (± 30 kHz offset, -20 dBm tone at input mixer) | ||||
HP 8591E, 8594E, 8595E, 8596E | < -65 dBc | |||
HP 8593E | < -65 dBc (applied freq ≤18 GHz); < -60 dBc (applied freq ≤22 GHz) | |||
Residual Responses (input terminated, 0 dB attenuation) | ||||
1 MHz to 1.8 GHz (HP 8591E Opt 001) | < - 38 dBmV | |||
150 kHz to 1.8 GHz (HP 8591E) | < -90 dBm | |||
150 kHz to 6.5 GHz (HP 8593E, 8595E, 8596E) | < -90 dBm | |||
Display Range | ||||
Log Scale | 0 to -70 dB from ref level is calibrated: 0.1, 0.2, 0.5 dB/div and 1 to 20 dB/div in 1 dB steps; 8 div displayed | |||
Linear Scale | 8 divisions | |||
Scale Units | dBm, dBmV, dBµV, V, Ω | |||
Marker Readout Resolution | ||||
Log Scale | 0.05 dB | |||
Linear Scale | 0.05% of ref level | |||
Fast Time Sweep for Zero Span (Opt 101 or 301, 20 µs to 20 ms) | ||||
≤1 GHz | 0.7% of ref level for linear scale | |||
> 1 GHz | 1.0% of ref level for linear scale | |||
Reference Level | ||||
Range | Same as amplitude range | |||
Resolution | ± 0.01 dB for log scale; ± 0.12% of ref level for linear scale | |||
Accuracy | ± 0.3 dB at -20
dBm 0 to -59.9 dBm: ± (0.3 dB + 0.01 X dB from -20 dBm) |
|||
Frequency Response (10 dB input attenuation) | ||||
Absolute (referenced to 300 MHz CAL OUT) | ||||
HP 8591E, 8594E | ± 1.5 dB | |||
HP 8595E | ± 1.5 to ± 2.0 dB | |||
HP 8596E | ± 1.5 to ± 2.5 dB | |||
HP 8593E | ± 1.5 to ± 5.0 dB (preselector peaked) | |||
Relative Flatness (referenced to midpoint between highest and lowest frequency response deviations) | ||||
HP 8591E, 8594E | ± 1.0 dB | |||
HP 8595E | ± 1.0 to ± 1.5 dB | |||
HP 8596E | ± 1.0 to ± 2.0 dB | |||
HP 8593E | ± 1.0 to ± 2.0 dB (preselector peaked) | |||
Calibrator Output Amplitude | -20 dBm ± 0.4 dB; + 28.75 dBmV ± 0.4 dB, HP 8591 Opt 001 | |||
Resolution Bandwidth Switching Uncertainty (ref to 3 kHz RBW, at ref level) | ||||
3 kHz to 3 MHz RBW | ± 0.4 dB | |||
1 kHz RBW | ± 0.5 dB | |||
30 Hz to 300 Hz RBW | ± 0.6 dB | |||
Log to Linear Switching | ± 0.25 dB at ref level | |||
Display Scale Fidelity | ||||
Log Incremental Accuracy (0 to -60 dB from ref level) | ± 0.4 dB/4 dB | |||
Log Maximum Cumulative (0 to -70 dB from ref level) | ||||
1 kHz to 3 MHz RBW | ± (0.3 + 0.01 X dB from ref level) | |||
30 to 300 Hz RBW | ± (0.4 + 0.01 X dB from ref level) | |||
Linear Accuracy | ± 3% of ref level | |||
GENERAL SPECIFICATIONS | ||||
Temperature | ||||
Operating | 0° to +55°C | |||
Storage | -40° to +75°C | |||
EMI Compatibility | Conducted and radiated interference CISPR Pub. 11 and Messempfaenger Postverfuegung 526/527/79 | |||
Audible Noise | < 37.5 dBA pressure and < 5.0 Bels power (ISODP7779) | |||
Power Requirements | ||||
On (line 1) | 90 to 132 V rms, 47
to 440 Hz 195 to 250 V rms, 47 to 66 Hz Power Consumption < 500 VA; < 180 Ω |
|||
Standby (line 0) | Power Consumption < 7 Ω | |||
User Program Memory (nominal) | 121 KB nonvolatile RAM | |||
Data Storage (nominal) | ||||
Internal | 50 traces; 8 states | |||
External | ||||
Memory Cards | HP 85700A (32 KB),
24 traces or 32 states HP 85702A (128 KB), 99 traces or 128 states |
|||
Video Cassette Recorder (VCR) | Continuous video recording of display supported through composite video output | |||
Size (nominal, without handle, feet, or cover) | 325 mm W X 163 mm H X 427 mm D | |||
Weight | 14.5 kg (HP 8591E): 16.4 kg (HP 8593E, 8594E, 8595E, 8596E) | |||
Option 010 AND 011 BUILT-IN TRACKING GENERATORS | ||||
Option 010 (50 Ω) is available for all HP 8590 Series spectrum analyzers except the HP 8592D. Opt 011 (75 Ω) is available for the HP 8590D and 8591E only. | ||||
Frequency Range | ||||
Opt 010 | 100 kHz to 1.8 GHz (HP 8590D, 8591E); 300 kHz to 2.9 GHz (HP 8593E, 8594E, 8595E, 8596E) | |||
Opt 011 | 1 MHz to 1.8 GHz (HP 8590D, 8591E) | |||
Output Level | ||||
Range | ||||
Opt 010 | 0 to -15 dBm (HP 8590D); 0 to -70 dBm (HP 8591E); -1 to -66 dBm (HP 8593E, 8594E, 8595E, 8596E) | |||
Opt 011 | +42.8 to -27.8 dBm V (HP 8590D); +42.8 to -27.2 dBmV (HP 8591E) | |||
Resolution | 0.1 dB | |||
Absolute Accuracy | ± 1.5 dB (HP 8590D, 8592D); ± 1.0 dB (HP 8591E); ± 0.75 dB (HP 8593E, 8594E, 8595E, 8596E) | |||
Vernier | ||||
Range | 15 dB (HP 8590D); 10 dB (HP 8591E); 8 dB (HP 8593E, 8594E, 8595E, 8596E) | |||
Accuracy | ± 1.0 dB (HP 8590D); ± 0.25 dB (HP 8591E); ± 0.8 dB (HP 8593E, 8594E, 8595E, 8596E) | |||
Output Flatness | ± 1.75 dB (HP 8590D, 8591E); ± 2.0 dB, > 10 MHz (HP 8593E, 8594E, 8595E, 8596E) | |||
Spurious Output | ||||
Harmonic Spurs | 0 dBm + 42.8 dBmV output, < -20 dBm (HP 8590D); < -25 dBc (HP 8593E, 8594E, 8595E, 8596E) | |||
Nonharmonic Spurs | < -30 dBc | |||
Dynamic Range (characteristic; max output level -TG feed-though) | ||||
Opt 010 | 106 dB (HP 8590D, 8591E); 108 dB (HP 8594E, > 400 kHz); 111 dB (HP 8595E, 8596E, > 400 kHz); 113 dB (HP 8593E, > 400 kHz) | |||
Opt 011 | 100 dB | |||
Power Sweep | ||||
Range | ||||
Option 010 | -15 dBm to 0 dBm (HP 8590D); -75 dBm to 0 dBm (HP 8591E); -66 dBm to -1 dBm in 8 dB increments (HP 8593E, 8594E, 8595E, 8596E) | |||
Option 011 | -27.8 dBmV to +42.8 dBmV (HP 8590D); -32.2 to +42.8 dBmV (HP 8591E) | |||
Resolution | 0.1 dB |
Specifications apply to either 8590D or 8592D analyzer unless otherwise noted. | ||
FREQUENCY | ||
Frequency Range | ||
HP 8590D | ||
50 Ω | 9 kHz to 1.8 GHz | |
75 Ω (Opt 001) | 1 MHz to 1.8 GHz | |
HP 8592D | 9 kHz to 22 GHz | |
HP 8592D (Opt 026) | 9 kHz to 26.5 GHz | |
Band | LO Harmonic = N | Center Frequency |
0 | 1 | 9 kHz to 2.9 GHz |
1 | 1 | 2.75 to 6.5 GHz |
2 | 2 | 6.0 to 12.8 GHz |
3 | 3 | 12.4 to 19.4 GHz |
4 | 4 | 19.1 to 22.0 GHz |
4 | 4 (Opt 026) | 19.1 to 26.5 GHz |
Frequency Reference (HP 8590D Opt 013) | ||
Aging | ± 2 X 10-6/year | |
Temperature Stability | ± 5 X 10-6 | |
Initial Achievable Accuracy (start, stop, center, marker) | ||
HP 8590D | ± (5 MHz + 1% of freq span) | |
HP 8590D Opt 013 | ± (freq readout X freq ref error + span accuracy + 1% of span + 20% of RBW + 100 Hz) | |
HP 8592D | ± [(5 X N) MHz + 0.01% of center freq + 2% of freq span] | |
Marker Count Accuracy (HP 8590D Opt 013) | ||
Span ≤10 MHz | ± (marker freq X freq ref error + counter resolution + 100 Hz) | |
Span > 10 MHz | ± (marker freq X freq ref error + counter resolution + 1 kHz) | |
Counter Resolution | Span ≤10 MHz, selectable from 10 Hz to 100 kHz; span > 10 MHz, selectable from 100 Hz to 100 kHz | |
Frequency Span | ||
Range | ||
HP 8590D | 0 Hz (zero span), 10 kHz to 1.8 GHz | |
HP 8592D | 0 Hz, [50 X N] kHz to 19.25 GHz | |
Resolution | Four digits | |
Accuracy | ± 3% of span | |
Sweep Time | ||
Range | 20 ms to 100 s | |
Accuracy | ± 3% | |
Sweep Trigger | Free run, single, line, video, external | |
Resolution Bandwidth (characteristic) | 1 kHz to 3 MHz (3 dB) in 1, 3, 10 sequence ± 20% accuracy; 9 kHz and 120 kHz (6 dB) EMI bandwidths | |
Video Bandwidth Range | 30 Hz to 1 MHz in 1, 3, 10 sequence | |
Stability | ||
Noise Sidebands (1 kHz RBW, 30 Hz VBW and sample detector) | ≤-95 dBc/Hz + 20 log N at > 30 kHz offset from CW signal | |
System-related Sidebands | ≤-65 dBc + 20 log N at > 30 kHz at offset from CW signal | |
Comb Generator Frequency (HP 8592D) | 100 MHz fundamental freq | |
Accuracy | ± 0.007% | |
AMPLITUDE | ||
Amplitude Range | ||
HP 8590D, 8592D | Displayed average noise level to +30 dBm | |
HP 8590D Opt 001 | Displayed average noise level to +75 dBmV | |
Maximum Safe Input Level (input attenuator 10 dB) | ||
Average Continuous Power | ||
HP 8590D, 8592D | +30 dBm (1Ω) | |
HP 8590D Opt 001 | +75 dBmV (0.4Ω) | |
Peak Pulse Power | ||
HP 8590D | +30 dBm (1Ω); +75 dBmV (0.4Ω) (Opt 001) | |
HP 8592D | +50 dBm (100Ω) for < 10 µs pulse width and < 1% duty cycle, input atten 30 dB | |
dc | ||
HP 8590D | 25 Vdc; 100 Vdc (Opt 001) | |
HP 8592D | 0 Vdc | |
Gain Compression (> 10 MHz) | ≤ 0.5 dB (total power at input mixer = -10 dBm) | |
Displayed Average Noise Level (input terminated, 0 dB atten, 1 kHz RBW, 30 Hz VBW) | ||
HP 8590D | ≤ -115 to ≤-113 dBm; ≤-63 to ≤-61 dB,V (Opt 001) | |
HP 8592D | ≤-112 to ≤-92 dBm; ≤-112 to ≤ -87 dBm (Opt 026) | |
Spurious Responses | ||
Second Harmonic Distortion ( >5 MHz) | ||
HP 8590D | < -70 dBc for -45 dBm tone at input mixer | |
HP 8592D | ||
10 MHz to 2.9 GHz | < -70 dBc for -40 dBm tone at input mixer | |
2.75 GHz | < -100 dBc for -10 dBm tone at input mixer (or below DANL) | |
Third-order Intermodulation | ||
HP 8590D | ||
Distortion > 5 MHz | < -70 dBc for two -30 dBm tones at input mixer and > 50 kHz separation | |
Other Input-related | < -65 dBc at 30 kHz offset, for -20 dBm tone at input mixer | |
HP 8592D | ||
Distortion > 10 MHz | < -70 dBc for two -30 dBm tones at input mixer and > 50 kHz separation | |
Other Input-related | < -65 dBc at 30 kHz offset, for -20 dBm tone at input mixer, ≤18 GHz; < -60 dBc for -20 dBm tone at input mixer, ≤22 GHz | |
Display Range | ||
Log Scale | 0 to -70 dB from ref level is calibrated; 0.1, 0.2, 0.5 dB/div and 1 to 20 dB/div in 1 dB steps; 8 div displayed | |
Linear Scale | 8 divisions | |
Scale Units | dBm, dBmV, dBµV, V, Ω | |
Marker Readout Resolution | 0.05 dB for log scale; 0.5% of reference level for linear | |
Reference Level | ||
Range | Same as amplitude range | |
Resolution | 0.01 dB for log scale; 0.12% of ref level for linear | |
Accuracy | ± 0.3 dB @ -20 dBm 0 dBm to -59.9 dBm: ± (0.3 dB + 0.01 X dB from -20 dBm) |
|
Frequency Response (10 dB input attenuation) | ||
Absolute (referenced to 300 MHz CAL OUT) | ||
HP 8590D | ± 1.5 dB | |
HP 8592D (preselector peaked in band > 0) | ± 1.5 to ± 5.0 dB | |
Relative | ± 1.0 dB, referred to midpoint between highest and lowest frequency response deviations | |
HP 8590D | ± 1.0 dB | |
HP 8592D (preselector peaked in band > 0) | ± 1.0 to ± 2.0 dB | |
HP 8590D Opt 001 | +28.75 dBmV ± 0.4 dB | |
Resolution Bandwidth Switching Uncertainty (ref to 3 kHz RBW, at ref level) | ± 0.4 dB for 3 kHz to 3 MHz RBW; ± 0.5 dB for 1 kHz | |
Log to Linear Switching | ± 0.25 dB at ref level | |
Display Scale Fidelity | ||
Log Incremental Accuracy | ± 0.4 dB/4 dB, 0 to -60 dB from ref level | |
Log Maximum Cumulative | ± (0.4 dB + 0.01 X dB from ref level), 0 to -70 dB from ref level | |
Linear Accuracy | ± 3% of ref level | |
GENERAL SPECIFICATIONS | ||
Same as for HP 8590E-Series. |