Measuring the cband cal values.
jan,2001 (last updated 02mar02)
Links to sections:
Measuring the hcorcal
Measuring the ratio of the "other"
cals to the hcorcal.
Links to plots:
The
hcorcal measured values:
The
ratio of the other cals to the hcorcal:
Comparing
Tsys measurements 2002 before, after cals updated.
Intro:
A 2 db pad was removed from Pol A of the cband system
(after the cal switches). The cal value was measured on 19jan01 using absorber
in front of the horn. On 22jan01 the cal value was measured while tracking
blank sky. The correlator was configured as 25 Mhz by 1024 lags by 4 sbc.
3950 to 6050 MHz was covered in 100 Mhz chunks (22 steps). At each
step a 5 second integration with the cal on was followed by a 5 second
integration with the cal off. The frequency range 3950 to 6050 was covered
3 times (each pass separated by about 6 minutes). On 20dec01 blank sky
was tracked and the ratio of the "other" cals was measured relative to
the hcorcal. see
measuring cals on the telescope using sky and absorber for how the
cal value was computed.
Measuring the hcorcal:
(..top)
The temperatures used for cband were:
Temperatures Used
Trcvr |
from amplifiers |
8 K |
Tomt |
orthmode transducer |
4 K |
Tabs |
absorber temperature |
301 K |
Tsky |
sky + scattered ground radiation |
3 + 15=18K |
Trcvr |
Tamp+Tomt |
8 + 4 = 12K |
The
hcorcal measured values
show the results of these measurements. Figure
1 plots the cal Values measured the 3 different ways:
Red shows cal polA, polB measured relative to the absorber.
Green shows cal polA,polB measured relative to the sky.
Light Blue is cal polA, polb measured using the ratios of the cals with
absorber and sky.
The black line is the cal values measured in 1999 prior to the 2db pad
being removed from polA.
All three methods give consistent values for 4300 through 5000 Mhz (of
course I twiddled the value of Tsky for this to happen!). Outside this
range the Cal values measured from Tsky do not agree with those from Tabs
or the ratios. For each line and frequency there are 3 separate measurements.
They lie on top of each other so there was no problem with time variations
of gain or Tsky changing with zenith angle. Between 4400 and 6000 Mhz the
old value of Calb (which shouldn't have changed varies from calAbs,calRatio
by about 2 degK.
Below 4200 Mhz the cal measured from the sky and
the absorber diverge in opposite directions. calAbs climbs rapidly (especially
polB) while calSky drops off. This is most likely explained by the OMT
being lossy in this region. Losses would cause the Tabs reaching the amps
to decrease (TempOmt < Tabs) hence i used a Tabs too large. Tsky reaching
the receiver would increase (tempOmt > Tsky) and i used a value of Tsky
too small.
The cal values from 4300 Mhz to 5000 Mhz is probably
ok. The values outside this range are suspect. The data also make you wonder
how well the receiver itself is going to work outside this limited range.
Figure 2 shows the narrow band resonances found while looking at the
absorber. It shows
(CalOn - CalOff)/(CalOff) * Tabs+Trcvr versus frequency for 3 bands.
Since Tabs is greater then omtTemp, the CalOff bandpass goes negative in
this region. When you do the division you get a positive going bump. Looking
at Calon - CalOff with no division, the difference is constant which shows
that the problem comes before the cal injection. The lowest resonance is
close to where the measure cal values start to diverge.
processing: x101/010122/cal.pro , x101/010119/doit.pro for resonances.
Measuring the
ratio of the "other" cals to the hcorcal. (..top)
On 20dec01 blank sky was tracked with the cb receiver.
A sequence of 10 cal on/off measurements were done using the correlator
configured as 25 MHz x 4 boards. The sequence was (hcorcal, hcal, hxcal,
h90cal, hcorcal, lcorcal, lcal, lxcal, l90cal, hcorcal). This sequence
was repeated covering 3950 to 6050 Mhz. The hcorcal was measured using
the sky and absorber (see above). The ratio of the "other" cals was then
computed relative to the hcorcal.(see
measuring cals on the telescope using sky and absorber ).
For each cal in the sequence, the cal on/off spectra
was computed and then cumfiltered to remove any "obvious" rfi. The ratio
with the hcorcal was then computed by interpolating between the 3 hcorcal
measurements. The interpolation was used since each cal is measured
relative to the calOff (or tsys) and tsys varies as we track blank sky.
The
ratio of the other cals to the hcorcal is shown in the plots:
-
Fig 1 top shows the high cals measured relative to tsys versus frequency.
The * are pol B. The vertical lines show the 100 Mhz steps. There were
4 25 Mhz measurements at each frequency setting. Some of the cal sizes
track each other since they come from the same diode (e.g.. polA hcorcal:hcal,
hxcal:h90cal, polB hcorcal,h90cal). The bottom plot has the low cals. The
bumps at 5750 and 6000 Mhz are rfi.
-
Fig 2 top plots the ratio of the cal/hcorcal for the high cals. The horizontal
red, black line have diode1->polA and diode 1->polB. This is the same configuration
for the hcorcal so the ratio is close to 1. The horizontal pink and bright
red lines have diode2->polA and diode2->polB. The ratio of about .88 is
the ratio of diode2/diode1. The brownish,green diagonal line has diode2->polA
through the 0deg leg of the 90 deg hybrid after diode2. This normally gives
the 90 deg phase shift cals. The slope in frequency is the response
of the hybrid.
-
Fig 2 bottom is the ratio of the low cals to the hcorcal. The added ripple
(relative to the hcals) is probably the response of the 10 db coupler used
to generate the lowcal. The cal values were computed by averaging the cal
values over each 100 Mhz step. The black lines with the * show these values.
-
Fig 3 show how the hcorcal measurement varied for the 3 samples taken at
each frequency. The 2nd (red) and 3rd (green) hcorcal measurement were
divided by the first hcorcal measurement (and then subtracted 1). Some
of the slope is the change in Tsys with za (decreases at the start then
increases at the end). 5800 and 6000 Mhz had rfi.
To check the cals, the Tsys
for jan,feb 2002 using the old and new cal values were plotted.
-
The top figure plots Tsys using the old cal values. The solid lines are
polA and the dashed lines are polB. The system temperature is all over
the place.
-
The bottom figure plots Tsys using the new cal values.
The scatter in polB is < .5K (1.5%). The Tsys for
polA are close except for hcal which differs by over 1K from the others.
The same cal value is used for polA hcorcal and hcal since it is diode
1 going to polA. The software assumes that the cal value for diode 1->polA
is independent of which diode is feeding polB. From looking at the Tsys
measurements it appears that this assumption is not always true. The "other"
cal values above 5800 Mhz were not computed because of rfi (the software
will use the value at 5800 Mhz).
processing: x101/Y01/011220/cbcal/cbcal.pro,chkcbcal.pro
home_~phil